OL 752-12U Standard of Spectral Irradiance (Seasoned)

The OL 752-12U Seasoned Lamp for the OL 752-12 Series Standard of Spectral Irradiance is calibrated for spectral irradiance over the wavelength range of 200 to 400 nm It uses a stable, 40-W deuterium lamp. When mounted in the plug-in/pre-aligned housing, the working distance is 10 cm. This short working distance generates considerably higher irradiance levels than that obtained with conventional ultraviolet irradiance standards.

The OL 752-12U uses the OL 46 Deuterium Lamp Power Supply.

GET A QUOTE

Description

The OL 752-12U Seasoned Lamp for the OL 752-12 Series Standard of Spectral Irradiance is calibrated for spectral irradiance over the wavelength range of 200 to 400 nm It uses a stable, 40-W deuterium lamp. When mounted in the plug-in/pre-aligned housing, the working distance is 10 cm. This short working distance generates considerably higher irradiance levels than that obtained with conventional ultraviolet irradiance standards. The OL 752-12 is calibrated relative to the NIST Ultraviolet Irradiance Standard for the region below 250 nm and to the NIST High-Accuracy Scale of Spectral Irradiance for the 250 to 400 nm region. The NIST Ultraviolet Irradiance Standard has an uncertainty that varies from ± 7.5% at 200 nm to ± 5% at 250 nm. The OL 752-12 has a transfer uncertainty relative to the NIST Scales that varies from ± 3% at 200 nm to ± 1.5% at 400 nm.

Deuterium <em>(30-W)</em>

500 milliamps DC

@ 250 nm: 6 x 10<sup>-8</sup> W/cm<sup>2</sup>nm <br> @ 550 nm: 3 x 10<sup>-5</sup> W/cm<sup>2</sup>nm <br> @ 1000 nm: 5 x 10<sup>-5</sup> W/cm<sup>2</sup>nm <br> @ 1600 nm: 3 x 10<sup>-5</sup> W/cm<sup>2</sup>nm

1.5 to 3%

≤ 0.06%/ hour <br> <em>*High stability option available</em>

Bulletin

OL 752-12 Plug-in Standards of Spectral Irradiance

Bulletin

OL 46D Deuterium Lamp Power Supply

OL 46D Deuterium Lamp Power Supply

The OL 46D features complex automatic lamp starting and operating supplies. When the supply is turned on, a starting voltage...

1 YEAR STANDARD WARRANTY

Optronic Laboratories warrants that all goods supplied will be of the kind described or in any specification and drawings approved by Optronic Laboratories and will be free from defects in material and workmanship for one (1) year from the date of purchase by the original Purchaser.  During this period, Optronic Laboratories will at its option either repair or replace any goods that are found to be defective in material or workmanship, provided the goods are returned to Optronic Laboratories  in Orlando, Florida, with all shipping, insurance, and delivery charge prepaid (DDU Orlando, Florida).  This Warranty does not extend to batteries, fuses, and glass phototubes (if any), or other items of limited durability, nor does this warranty cover damage to goods caused by leaky or otherwise defective batteries, by Purchaser use of improper batteries and by misuse or abuse, unauthorized alteration, excessive line voltage, excessive temperatures (above 160 degrees F), extreme environmental conditions (such as extremely dusty or wet environment), corrosive atmospheres or servicing by unauthorized personnel.  The items returned shall only be accepted when accompanied by a written statement setting forth the nature and suspected cause of the alleged deficiencies.  After the initial one (1) year warranty, repairs performed by Optronic Laboratories are covered under a limited ninety (90) day warranty. The warranty repaired (and/or replacement) items will be returned DDU to customer. THERE ARE NO OTHER WARRANTIES, EXPRESS OR IMPLIED, (INCLUDING THE WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE OR IMPLIED WARRANTY OF MERCHANTABILITY) OTHER THAN THE WARRANTY SET FORTH HEREIN.